GPIB Command Dictionary

CONFigure subsystem

:OFLevel:ANTenna <real number> :OFLevel:ANTenna?

These commands set/query the path loss from the device-under-test to the Test Set’s ANT IN port. This correction is applied when the CONF:OFL:MODE ‘On’ command is used.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

CONF:OFL:ANT <real number> !values: 100.0 to 100.0

CONF:OFL:ANT?

Screen/field equivalent

CONF:OFL:ANT controls the Antenna In field of the INSTRUMENT CONFIGURE screen. This value in this field is used when the RF Level Offset field is set to On.

:OFLevel:DUPLex <real number> :OFLevel:DUPLex?

These commands set/query the path loss from the DUPLEX OUT connector to the device-under-test. This correction is applied when the CONF:OFL:MODE ‘On’ command is used.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

CONF:OFL:DUPL <real number> !values: 100.0 to 100.0

CONF:OFL:DUPL?

Screen/field equivalent

CONF:OFL:DUPL controls the Duplex Out field on the INSTRUMENT CONFIGURE screen. This value in this field is used when the RF Level Offset field is set to On.