Agilent Technologies E6380-90073 RFGenerator string RFGenerator?, RLEVel real number RLEVel?

Models: E6380-90073

1 352
Download 352 pages 53.27 Kb
Page 330
Image 330

GPIB Command Dictionary

SANalyzer subsystem

:RFGenerator <string> :RFGenerator?

This command selects between the fixed frequency generator and the tracking generator. The query form returns the mode setting of the generator.

Syntax

SAN:RFG ‘Track’

‘Fixed’

SAN:RFG?

Screen/field equivalent

SAN:RFG controls the Track/Fixed field on the RF Gen menu of the SPEC ANL screen.

:RLEVel <real number> :RLEVel?

These commands set/query the reference level of the spectrum analyzer.

This command utilizes the “Real Number Setting Syntax” on page 299, but does not use the :STATe command.

Syntax

SAN:RLEV <real number>

SAN:RLEV?

Screen/field equivalent

SAN:RLEV controls the Ref Level field on Main menu of the SPEC ANL screen.