1 352
Download 352 pages 53.27 Kb
Page 133
Image 133

GPIB Command Dictionary

CDANalyzer subsystem

:IS2000:GAIN[:SETTing] <string> :IS2000:GAIN[:SETTing]?

For IS-2000 only. These commands set/query the gain of the Code Domain analyzer. This command will only be enabled when autoranging is turned off (CDAN:IS2000:GAIN:MODE ‘Hold’).

This command may conflict with the CDMA analyzer’s settings CAN:PATH:GAIN and CAN:PATH:MODE. For example, if autoranging is turned off (‘Hold’) in the CDMA analyzer and turned on in the Code Domain Analyzer (‘Auto’), the ‘Hold’ setting will always override the ‘Auto’ setting. When autoranging is turned off, setting either analyzer’s gain will automatically change the other’s to the same setting.

Use the CAN:POW:GAIN command when making an average power measurement.

Syntax

CDAN:IS2000:GAIN[:SETT] ‘0 dB’ ‘6 dB’ ‘12 dB’ ‘18 dB’ ‘24 dB’ ‘30 dB’ ‘36 dB’

CDAN:IS2000:GAIN[:SETT]?

Screen/field equivalent

CDAN:IS2000:GAIN[:SETT] sets the lower subfield of the Gain image of the CODE DOMAIN screen. This subfield can only be set when the upper subfield is set to Hold when the CDMA standard is set to “IS-2000”.

Chapter 1

133