GPIB Command Dictionary

CDANalyzer subsystem

:MODE <string> :MODE?

These commands set/query the measurement state of the code domain analyzer.

CDAN:MODE can be overridden with the TRIG:MODE:RETR REP and TRIG:MODE:RETR SING trigger commands. When the trigger commands are used, the state of the field on the display is not updated. (For instance, the screen may show Single selected when TRIG:MODE:RETR REP is used.)

Syntax

CDAN:MODE ‘Single’

‘Cont’

CDAN:MODE?

Screen/field equivalent

CDAN:MODE controls the Single/Cont subfield of the Measurement field on the Main menu of the CODE DOM screen, when the CDMA standard is set to “IS-95 Only”.

:PATH:GAIN <real number> :PATH:GAIN?

These commands set/query the gain into the code domain analyzer. This command fine tunes the gain setting. This value is only selectable through GPIB.

See also, CAN:PATH:GAIN on page 106.

Syntax

CDAN:PATH:GAIN <real number> !values from 0 to 36 dB in 0.1 dB steps

CDAN:PATH:GAIN?

Screen/field equivalent

CDAN:PATH:GAIN has no equivalent field. This applies when the

CDMA standard is set to “IS-95 Only”.

Chapter 1

125

Page 125
Image 125
Agilent Technologies E6380-90073 manual Mode string MODE?, Pathgain real number PATHGAIN?