Commands Listed in Alphabetical Order
Examples BUS:B1:FLEXRAY:CHANNEL B sets the FLEXRAY id format to channel B.
BUS:B1:FLEXRAY:CHANNEL? might return BUS:B1:FLEXRAY:CHANNEL A
indicating that FLEXRAYid format is channel A.
BUS:B<x>:FLEXray:SIGnal
Sets or returns the FLEXRAY standard.
Group Bus
Syntax BUS:B<x>:FLEXray:SIGnal {BDIFFBP|BM|TXRX}
BUS:B<x>:FLEXray:SIGnal?
Arguments BDIFFBP sets the FLEXRAY standard to BDIFFBP.
BM sets the FLEXRAY standard to BM.
TXRX sets the FLEXRAY standard to TXRX.
Examples BUS:B1:FLEXRAY:SIGNAL BM sets the FLEXRAY standard is BM.
BUS:B1:FLEXRAY:SIGNAL? might return BUS:B1:FLEXRAY:SIGNAL
BDIFFBP indicating the FLEXRAY standard is BDIFFBP.
BUS:B<x>:FLEXray:SOUrce
Sets or returns the FLEXRAY data source.
Group Bus
Syntax BUS:B<x>:FLEXray:SOUrce
{CH1|CH2|CH3|CH4|D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|D
1
BUS:B<x>:FLEXray:SOUrce?
Arguments CH<x> sets the FLEXRAYsource to channel x, where x is 1 to 4.
D<x> sets the FLEXRAYsource to digital channel x, where x is 1 to 16.
MSO4000 and DPO4000 Series Programmer Manual 2-95