Commands Listed in Alphabetical Order
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimitSets or returns the maximum time that the selected pattern may be true and still
generate an A logic pattern trigger.
Group Trigger
Syntax TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit <NR3>
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit?
Arguments <NR3> specifies the maximum amount of time to hold the pattern true.
Examples TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT 10.0E+00 sets the
maximum time that the selected pattern may hold true (and generate an A logic
pattern trigger) to 10 s.
TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT? might return
TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT 8.0000E-9 indicating that
the selected pattern may hold true for up to 8 ns and still generate an A logic
pattern trigger.
TRIGger:A:LOGIc:PATtern:WHEn:MORELimitSets or returns the minimum time that the selected pattern may be true and still
generate an A logic pattern trigger.
Group Trigger
Syntax TRIGger:A:LOGIc:PATtern:WHEn:MORELimit <NR3>
TRIGger:A:LOGIc:PATtern:WHEn:MORELimit?
Arguments <NR3> specifies the minimum amount of time to hold the pattern true.
Examples TRIGGER:A:LOGIC:PATTERN:WHEN:MORELIMIT 10.0E+00 sets the minimum
time that the selected pattern may hold true (and generate an A logic pattern
trigger) to 10 s.
TRIGGER:A:LOGIC:PATTERN:WHEN:MORELIMIT? might return
TRIGGER:A:LOGIC:PATTERN:WHEN:MORELIMIT 8.0000E-9 indicating that
the selected pattern must hold true for at least 8 ns to generate an A logic pattern
trigger.
2-436 MSO4000 and DPO4000 Series Programmer Manual