Commands Listed in Alphabetical Order
Related Commands TRIGger:B:EVENTS?
Arguments <NR1> is the number of B trigger events, which can range from 1 to 10,000,000.
Examples TRIGGER:B:EVENTS:COUNT 4 specifies that the B trigger will occur four trigger
events after the A trigger.
TRIGGER:B:EVENTS:COUNT? might return :TRIGGER:B:EVENTS:COUNT 2
indicating that two events must occur after the A trigger before the B trigger
can occur.
TRIGger:B:LEVelSets or returns the level for the B trigger.
Group Trigger
Syntax TRIGger:B:LEVel {TTL|<NR3>}
TRIGger:B:LEVel?
Related Commands TRIGger:A:LEVel,TRIGger:B,TRIGger:B:EDGE:SOUrce
Arguments TTL specifies a preset TTL high level of 1.4 V.
<NR3> is the B trigger level, in volts.
Examples TRIGGER:B:LEVEL TTL sets the B trigger level to 1.4 V.
TRIGGER:B:LEVEL? might return :TRIGGER:B:LEVEL 173.0000E-03
indicating that the B trigger level is currently set at 173 mV.
TRIGger:B:LEVel:CH<x>Sets or returns the B trigger level for channel <x>, where x is the channel number.
Each Channel can have an independent Level.
Group Trigger
Syntax TRIGger:B:LEVel:CH<x> {ECL|TTL|<NR3>}
TRIGger:B:LEVel:CH<x>?
MSO4000 and DPO4000 Series Programmer Manual 2-467