Commands Listed in Alphabetical Order
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:INDICators {OFF|ON|0|1}
POWer:INDICators?
Arguments OFF or 0 turns off the measurement indicators.
ON or 1 turns on the measurement indicators.
Examples POWER:INDICATORS 1 turns on the indicator.
POWER:INDICATORS? might return :POWER:INDICators 0 indicating that
the indicator is off.
POWer:MODulation:SOUrce
Sets or returns the source waveform for modulation tests. The voltage source
waveform is specied using the POWer:VOLTAGESOurce command and the
current waveform is specied using the POWer:CURRENTSOurce command.
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:MODulation:SOUrce {VOLTage|CURRent}
POWer:MODulation:SOUrce?
Arguments VOLTage species voltage source waveform for modulation tests.
CURRent species current source waveform for modulation tests.
Examples POWER:MODULATION:SOURCE CURRent sets the modulation source to Current.
POWER:MODULATION:SOURCE? might return :POWER:MODULATION:SOURCE
VOLT indicating that the modulation source is set to Voltage.
MSO4000 and DPO4000 Series Programmer Manual 2-269