Commands Listed in Alphabetical Order
attached probe type. Only certain VPI probes support this command. See the
probe documentation for how to set these string arguments.
Group Vertical
Syntax CH<x>:PRObe:COMMAND <QString>, <QString>
Arguments <QString> are quoted strings specifying the probe command and value to set in
the probe attached to the specied channel.
Examples CH1:PROBE:COMMAND “MODE”, “4–4V1MHz” sets a Tektronix VPI-DPG probe
to the 4-4V1MHz mode.
CH1:PROBE:COMMAND “OUTPUT”, “ON” turns the output of a Tektronix
VPI-DPG probe on.
CH1:PROBE:COMMAND?“MODE” mightreturnCH1:PROBE:COM MAND
“MODE”,“4-4V1MHZ”.
CH<x>:PRObe:DEGAUss (No Query Form)
Starts a degauss auto-zero cycle on a TekVPI current probe attached to the input
channel specied by <x>, where x is the channel number.
Group Vertical
Syntax CH<x>:PRObe:DEGAUss EXECute
Arguments EXECute initiates the degauss operation.
CH<x>:PRObe:DEGAUss:STATE? (Query Only)
Returns the state of the probe degauss for the channel specied by <x>, where is x
is the channel number.
NOTE. This command will return
PASSED
for probes that do not support degauss
operations.
Group Vertical
2-124 MSO4000 and DPO4000 Series Programmer Manual