Commands Listed in Alphabetical Order
FACtory (No Query Form)
Resets the oscilloscope to its factory default settings.
This command does the following:
Clears the Event Status Enable Register
Clears the Service Request Enable Register
Sets the Device Event Status Enable Register to 255
Purges all dened aliases
Enables all Command Headers
Sets the macro denedby*DDTtoa"zero-lengtheld"
Clears the pending operation ag and associated operations
This command does not reset the following:
Communication settings
Selected GPIB address.
Sate of the VXI-11 (Ethernet IEEE Std 488.2) interface.
Calibration data that affects device specications
Protected user data
Stored settings
Power On Status Clear Flag
Oscilloscope password
Group Save and Recall
Syntax FACtory
Related Commands *PSC,*RCL,RECAll:SETUp,*RST,*SAV,SAVe:SETUp
Arguments None
Examples FACTORY resets the oscilloscope to its factory default settings.
2-172 MSO4000 and DPO4000 Series Programmer Manual