Commands Listed in Alphabetical Order
TRIGger:A:EDGE:SOUrce?
Related Commands TRIGger:A:EDGE:SLOpe,TRIGger:A:EDGE:COUPling
Arguments CH1–CH4 species an analog input channel as the A edge trigger source.
D0–D15 species a digital channel as the source (MSO models only).
EXT species an external trigger using the Aux In connector located on the front
panel of the oscilloscope.
LINE species the AC line as the trigger source.
AUX species the Auxiliary Input as the trigger source (if available on your
oscilloscope).
Examples TRIGGER:A:EDGE:SOURCE CH1 sets channel 1 as the A edge trigger source.
TRIGGER:A:EDGE:SOURCE? might return :TRIGGER:A:EDGE:SOURCE CH1
indicating that channel 1 is the A edge trigger source.
TRIGger:A:HOLDoff? (Query Only)
Returns the A trigger holdoff parameters. These parameters specify the time
period during which the trigger circuitry is not looking to generate a trigger event.
Group Trigger
Syntax TRIGger:A:HOLDoff?
Related Commands TRIGger:A:HOLDoff:TIMe
Examples TRIGGER:A:HOLDOFF? might return :TRIGGER:A:HOLDOFF:TIME
900.0000E-09;BY DEFAULT indicating that the A edge trigger holdoff time
(by default) is 900 ns.
TRIGger:A:HOLDoff:TIMe
Sets or returns the A trigger holdoff time.
Group Trigger
MSO4000 and DPO4000 Series Programmer Manual 2-427