Commands Listed in Alphabetical Order
Related Commands TRIGger:A:RUNT:WIDt h
Arguments OCCURS argument species a trigger event if a runt of any detectable width occurs.
LESSthan argument sets the oscilloscope to trigger if the a runt pulse is detected
with width less than the time set by the TRIGger:A:RUNT:WIDth command.
than argument sets the oscilloscope to trigger if the a runt pulse is detected with
width than the time set by the TRIGger:A:RUNT:WIDth command.
EQUal argument sets the oscilloscope to trigger if a runt pulse is detected with
width equal to the time period specied in TRIGger:A:RUNT:WIDth within
a ±5% tolerance.
NOTEQual argument sets the oscilloscope to trigger if a runt pulse is detected with
width greater than or less than (but not equal to) the time period specied in
TRIGger:A:RUNT:WIDth within a ±5% tolerance.
Examples TRIGGER:A:RUNT:WHEN THAN sets the runt trigger to occur when the
oscilloscope detects a runt in a pulse wider than the specied width.
TRIGGER:A:RUNT:WHEN? might return :TRIGGER:A:PULSE:RUNT:WHEN
OCCURS indicating that a runt trigger will occur if the oscilloscope detects a runt
of any detectable width.
TRIGger:A:RUNT:WIDth
Sets or returns the width for a runt trigger.
Group Trigger
Syntax TRIGger:A:RUNT:WIDth <NR3>
TRIGger:A:RUNT:WIDth?
Related Commands TRIGger:A:RUNT:WHEn
Arguments <NR3> species the minimum width, in seconds.
Examples TRIGGER:A:RUNT:WIDTH 15E-6 sets the minimum width of the pulse runt
triggerto15µs.
TRIGGER:A:RUNT:WIDTH? might return :TRIGGER:A:PULSE:RUNT:WIDTH
2.0000E-09 indicating that the minimum width of a pulse runt trigger is 2 ns.
MSO4000 and DPO4000 Series Programmer Manual 2-445