Commands Listed in Alphabetical Order
SETUP<x>:TIME? (Query Only)Returns the time when the oscilloscope setup was saved for the specified channel
<x>.
Group Save and Recall
Syntax SETUP<x>:TIME?
Examples SETUP2:TIME? might return “SETUP2:TIME: 15:24:07 which is the setup
time for channel 2.
*SREThe *SRE (Service Request Enable) command sets or returns the bits in the
Service Request Enable Register. For information, refer to Registers.
Group Status and Error
Syntax *SRE <NR1>
*SRE?
Related Commands *CLS,DESE,*ESE,*ESR?,EVENT?,EVMsg?,FACt ory ,*STB?
Arguments <NR1> is a value in the range from 0 through 255. The binary bits of the SRER
are set according to this value. Using an out-of-range value causes an execution
error. The power-on default for SRER is 0 if *PSC is 1. If *PSC is 0, the SRER
maintains the previous power cycle value through the current power cycle.
Examples *SRE 48 sets the bits in the SRER to binary 00110000.
*SRE? might return 32, showing that the bits in the SRER have the binary value
of 00100000.
*STB? (Query Only)*STB? (Read Status Byte) returns the contents of the Status Byte Register (SBR)
using the Master Summary Status (MSS) bit. For information, refer to Registers.
MSO4000 and DPO4000 Series Programmer Manual 2-389