Commands Listed in Alphabetical Order
MEASUrement:INDICators:STATE
Sets or returns the state of visible measurement indicators.
Group Measurement
Syntax MEASUrement:INDICators:STATE {OFF|MEAS<x>}
MEASUrement:INDICators:STATE?
Arguments OFF turns the visible measurement indicators off.
MEAS<x> displays the visible measurement indicators for measurement <x>,
where<x>canbe1,2,3,or4.
NOTE. Theremustbeanactivemeasurementbefore you can activate an indicator
for a specied measurement.
Examples MEASUREMENT:INDICATORS:STATE MEAS2 turns on the display of visible
measurement indicators for measurement 2.
MEASUREMENT:INDICATORS:STATE? might return
MEASUREMENT:INDICATORS:STATE OFF indicating that no measurement
indicators are active.
MEASUrement:INDICators:VERT<x>? (Query Only)
Returns the value of the specied vertical measurement indicator <x> from the
trigger point, where <x> can be 1, 2, 3, or 4. A negative value means that the
indicator is positioned earlier in the waveform record than the trigger point.
Group Measurement
Syntax MEASUrement:INDICators:VERT<x>?
Examples MEASUREMENT:INDICATORS:VERT2? might return
MEASUREMENT:INDICATORS:VERT2 -3.724507E-6 indicating that the
second measurement indicator is positioned 3.72 μs before the trigger point.
2-216 MSO4000 and DPO4000 Series Programmer Manual