Commands Listed in Alphabetical Order
Examples POWER:HARMONICS:IEC:FILTER OFF disables ltering of IEC harmonics.
POWER:HARMONICS:IEC:FILTER? might return
:POWER:HARMONICS:IEC:FILTER 1 indicating that the IEC harmonics are
ltered.
POWer:HARMonics:IEC:FUNDamental
Sets or returns the rated fundamental current for IEC harmonics. Valid values
ranges from 0 to 16 in increments of 0.1. The unit is ampere.
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:HARMonics:IEC:FUNDamental <NR3>
POWer:HARMonics:IEC:FUNDamental?
Arguments <NR3> is the fundamental current in amperes.
Examples POWER:HARMONICS:IEC:FUNDAMENTAL 1 sets the fundamental current to
1.0000 A.
POWER:HARMONICS:IEC:FUNDAMENTAL? might return
:POWER:HARMONICS:IEC:FUNDAMENTAL 16 indicating that the
fundamental current for IEC harmonics is set to 16.0000 A.
POWer:HARMonics:IEC:GROUPing
Sets or returns the enabled state for grouping of IEC harmonics.
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:HARMonics:IEC:GROUPing {OFF|ON|1|0}
POWer:HARMonics:IEC:GROUPing?