Commands Listed in Alphabetical Order
ROM runs just the IO board diagnostic group.
ACQ runs just the acquisition system diagnostic group.
DIAg:SELect:<function> (No Query Form)
Runs self-tests on the specied system subsystem.
Group Calibration and Diagnostic
Syntax DIAg:SELect:<function>
Arguments <function> species a single oscilloscope subsystem on which to run self tests
(diagnostics). Valid values are:
ACQ tests the acquisition system.
APPKey tests the application keys.
CPU tests the CPU.
DISplay tests the display.
FPAnel tests the front panel controls.
IO tests the IO ports.
ROM tests the system read only memory.
Examples DIAG:SELECT:CPU sets the oscilloscope to run just CPU tests.
DIAg:STATE (No Query Form)
This command starts or stops the oscilloscope self-test. Depending on the
argument, self-test capabilities are either turned on or off.
Group Calibration and Diagnostic
Syntax DIAg:STATE {EXECute|ABORt}
Arguments EXECute starts diagnostics.
ABORt stops diagnostics at the end of the current loop.
MSO4000 and DPO4000 Series Programmer Manual 2-157