Commands Listed in Alphabetical Order
*OPC? might return 1 to indicate that all pending OPC operations are finished.
PASSWord(No Query Form)Enables the *PUD and NEWpass set commands. Sending PASSWordwithout any
arguments disables these same commands. Once the password is successfully
entered, the *PUD and NEWpass commands are enabled until the oscilloscope
is powered off, or until the FACtory command or the PASSWord command with
no arguments is issued.
To change the password, you must first enter the valid password with the
PASSWordcommand and then change to your new password with the NEWpass
command. Remember that the password is case sensitive.
Group Miscellaneous
Syntax PASSWord <QString>
Related Commands NEWpass,*PUD
Arguments <QString> is the password and can includeupto10characters.Thefactory
default password is “XYZZY” and is always valid.
Examples PASSWORD “XYZZY” enables the *PUD and NEWpass set commands.
PASSWORD disables the *PUB and NEWpass set commands. You can still use the
query version of *PUD.
PICTBridge:DATEPrintEnables or disables printing the date on the print output
Group PictBridge
Syntax PICTBridge:DATEPrint {DEFLT|OFF|ON}
Arguments DEFLT is the default setting of the printer.
ON enables the date print on the print output.
OFF disables the date print on the print output.