Commands Listed in Alphabetical Order
CH<x>:YUNits
Sets or returns the units for the channel specied by <x>, where x is the channel
number. String arguments are case insensitive and any unsupported units will
generate an error.
Supported units are:
%, /Hz, A, A/A, A/V,A/W, A/dB, A/s, AA, AW,AdB, As, B, Hz, IRE, S/s, V,
V/A, V/V,V/W, V/dB, V/s, VV,VW, VdB, Volts, Vs, W,W/A, W/V, W/W,W/dB,
W/s,WA,WV,WW,WdB,Ws,dB,dB/A,dB/V,dB/W,dB/dB,dBA,dBV,dBW,
dBdB, day, degrees, div,hr, min, ohms, percent, s
Group Vertical
Syntax CH<x>:YUNits <QString>
CH<x>:YUNits?
Arguments QString is a string of text surrounded by quotes, specifying the supported units.
CLEARMenu (No Query Form)
Clears the current menu from the display. This command is equivalent to pressing
the front panel Menu off.
Group Miscellaneous
Syntax CLEARMenu
*CLS (No Query Form)
Clears the following:
Event Queue
Standard Event Status Register
Status Byte Register (except the MAVbit)
If the *CLS command immediately follows an <EOI>, the Output Queue and
MAVbit (Status Byte Register bit 4) are also cleared. The MAV bit indicates that
information is in the output queue. The device clear (DCL) GPIB control message
will clear the output queue and thus MAV. *CLS does not clear the output queue
or MAV.
MSO4000 and DPO4000 Series Programmer Manual 2-131