Commands Listed in Alphabetical Order
Examples POWER:SOA:MASK:DEFINE 10 species the SOA mask coordinates as
0.0E+0,0.0E+0,0.0E+0,0.0E+0,0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,300.0000,2.
4
POWER:SOA:MASK:DEFINE ? might return :POWER:SOA:MASK:DEFINE
0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,300.0000,0.0E+0
indicating that the SOA mask coordinates are set to
0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,300.0000,0.0E+0.
POWer:SOA:MASK:MAXAmps
Sets or returns the maximum current applied to SOA mask testing.
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:SOA:MASK:MAXAmps <NR3>
POWer:SOA:MASK:MAXAmps?
Arguments <NR3> is the maximum current applied to SOA mask testing.
Examples POWER:SOA:MASK:MAXAMPS 40 sets the SOA mask testing current to 40 A.
POWER:SOA:MASK:MAXAMPS ? might return :POWER:SOA:MASK:MAXAMPS
30.0000 indicating that the SOA mask testing maximum current is set to 30 A.
POWer:SOA:MASK:MAXVolts
Sets or returns the maximum voltage applied to SOA mask testing.
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:SOA:MASK:MAXVolts <NR3>
POWer:SOA:MASK:MAXVolts?
Arguments <NR3> is the maximum voltage applied to SOA mask testing.
2-292 MSO4000 and DPO4000 Series Programmer Manual