Commands Listed in Alphabetical Order
This value is in units of WFMOutpre:XUNit?. The query command will time
out and an error will be generated if the waveform specied by DATa:SOUrce
is not turned on.
Group Waveform Transfer
Syntax WFMOutpre:XZEro?
Related Commands DATa:SOUrce,WFMOutpre:XUNit?
Examples WFMOUTPRE:XZERO? might return :WFMOUTPRE:XZERO 5.6300E-9 indicating
that the trigger occurred 5.63 ns before the rst sample in the waveform record.
WFMOutpre:YMUlt? (Query Only)
Returns the vertical scale factor per digitizing level in units specied by
WFMOutpre:YUNit for the waveform specied by the DATa:SOUrcecommand.
The query command will time out and an error is generated if the waveform
specied by DATa:SOUrce is not turned on.
See the description of WFMInpre:YMUlt to see how this scale factor is used to
convert waveform sample values to volts.
Group Waveform Transfer
Syntax WFMOutpre:YMUlt?
Related Commands DATa:SOUrce,WFMInpre:YMUlt
Examples WFMOUTPRE:YMULT? might return :WFMOUTPRE:YMULT 4.0000E-3 indicating
that the vertical scale for the corresponding waveform is 100 mV/div (for 8-bit
waveform data).
WFMOutpre:YOFf? (Query Only)
Returns the vertical position in digitizing levels for the waveform specied by the
DATa:SOUrcecommand. The query command will time out and an error will be
generated if the waveform specied by DATa:SOUrceis not turned on.
MSO4000 and DPO4000 Series Programmer Manual 2-495