Commands Listed in Alphabetical Order
MEASUrement:REFLevel:ABSolute:MID[1]
Sets or returns the mid reference level, and is the 50% reference level when
MEASUrement:REFLevel:METHod is set to Absolute. This command affects the
results of period, frequency, delay,and all cyclic measurements.
NOTE. this command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group Measurement
Syntax MEASUrement:REFLevel:ABSolute:MID[1] <NR3>
MEASUrement:REFLevel:ABSolute:MID[1]?
Related Commands MEASUrement:REFLevel:METHod
Arguments <NR3> is the mid reference level, in volts. The default is 0.0 V.
Examples MEASUREMENT:REFLEVEL:ABSOLUTE:MID 1 .71 sets the mid reference level
to .71 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:MID? might return
:MEASUREMENT:REFLEVEL:ABSOLUTE:MID 0.7100E+00 indicating
that the absolute mid1 reference level is set to .71 V.
MEASUrement:REFLevel:ABSolute:MID2
Sets or returns the mid reference level for the "to" waveform when
taking a delay measurement, and is the 50% reference level when
MEASUrement:REFLevel:METHod is set to Absolute. This command affects the
results of delay measurements.
NOTE. this command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group Measurement
Syntax MEASUrement:REFLevel:ABSolute:MID2 <NR3>
MEASUrement:REFLevel:ABSolute:MID2?
MSO4000 and DPO4000 Series Programmer Manual 2-229