Commands Listed in Alphabetical Order
Group Power
Syntax POWer:REFLevel:PERCent:LOW <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:LOW?
Arguments <NR3> is the low value in percentage.
Examples POWER:REFLEVEL:PERCENT:LOW 15 sets the low reference level to 15% of the
signal amplitude.
POWER:REFLEVEL:PERCENT:LOW? might return
:POWER:REFLEVEL:PERCENT:LOW 10 indicating that the percentage low
reference level is set to 90% of the signal amplitude.
POWer:REFLevel:PERCent:MID<1-3>
Sets or returns the mid reference percent level for waveformmeasurements. Mid3
is specic to the power application.
Conditions This command requires a DPO4PWR application module.
Group Power
Syntax POWer:REFLevel:PERCent:MID<1-3> <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:MID<1-3>?
Arguments <NR3> is the mid value in percentage.
Examples POWER:REFLEVEL:PERCENT:MID1 25 sets the mid reference voltage to 25%.
POWER:REFLEVEL:PERCENT:MID1? might return
POWER:REFLEVEL:PERCENT:MID1 50.0000 indicating that the mid reference
voltage level is set to 50%.
POWer:RIPPle (No Query Form)
Does a vertical autoset for ripple measurements or sets the vertical offset to 0.
Conditions This command requires a DPO4PWR application module.