Commands Listed in Alphabetical Order
Group Status and Error
Syntax *RST
Related Commands FACt ory ,RECAll:SETUp,SAVe:SETUp,*PSC,*RCL,*SAV
Arguments None
Examples *RST resets the oscilloscope settings to factory defaults.
*SAV (No Query Form)
Stores the state of the oscilloscope to a specied memory location. You can use
the *RCL command to restore the oscilloscope to this saved state at a later time.
Group Save and Recall
Syntax *SAV <NR1>
Related Commands *RCL,RECAll:SETUp,SAVe:SETUp
Arguments <NR1> species a location in which to save the state of the oscilloscope. Location
values range from 1 through 10. Using an out-of-range location value causes an
execution error. Any settings that have been stored previously at this location
will be overwritten.
Examples *SAV 2 saves the current oscilloscope state in memory location 2.
SAVe:ASSIgn:TYPe
Sets or returns the assignment of the data to be saved when the front-panel Save
button is pressed.
Group Save and Recall
Syntax SAVe:ASSIgn:TYPe {IMAGe|WAVEform|SETUp}
SAVe:ASSIgn:TYPe?
MSO4000 and DPO4000 Series Programmer Manual 2-323