Commands Listed in Alphabetical Order

ACQuire? (Query Only)

Returns the following current acquisition parameters:
Stop after
Acquisition state
Mode
Number of averages
Sampling mode
Group Acquisition
Syntax ACQuire?
Related Commands ACQuire:MODe,ACQuire:NUMACq?,ACQuire:NUMAVg,
ACQuire:STOPAfter

ACQuire:MAGnivu

Sets or returns the MagniVu feature, which provides upto 32 times signal detail
for fast viewing of short events. This feature is not recommended for slow data
formats such as RS-232.
NOTE. MagniVuchannel sampling is available on MSO oscilloscopes only.
Group Acquisition
Syntax ACQuire:MAGnivu {<NR1>|OFF|ON}
ACQuire:MAGnivu?
Arguments <NR1> = 0 disables the MagniVu feature; any other value turns this feature on.
ON enables the MagniVu feature.
OFF disables the MagniVu feature.
MSO4000 and DPO4000 Series Programmer Manual 2-69