Commands Listed in Alphabetical Order
Syntax *PSC {OFF|ON|NR1>}
*PSC?
Related Commands DESE,*ESE,FACt ory ,*RST,*SRE
Arguments OFF sets the power-on status clear flag to false.
ON sets the power-on status clear flag to true.
<NR1> = 0 sets the power-on status clear flag to false. This disables the power-on
clear allowing the oscilloscope to possibly assert SRQ after power-on; any other
value sets the power-on status clear flag to true, enabling the power-onstatus clear
preventing any SRQ assertion after power on.
Examples *PSC 0 sets the power-on status clear flag to false.
*PSC? might return 1 to indicate that the power-on status clear flag is set to true.
*PUDSets or returns a string of Protected User Data. This data is protected by the
PASSWordcommand. You can modify it only by first entering the correct
password. This password is not necessary to query the data.
Group Status and Error
Syntax *PUD {<Block>|<QString>}
*PUD?
Related Commands PASSWord
Arguments <Block> is a block containing up to 300 ASCII characters.
<QString> is a string containing up to 300 ASCII characters.
Examples *PUD #229This oscilloscope belongs to me storesthe string "This
oscilloscope belongs to me" in the user protected data area.
*PUD? might return #221PROPERTY OF COMPANY X
2-316 MSO4000 and DPO4000 Series Programmer Manual