Commands Listed in Alphabetical Order
Examples TRIGGER:A:LOWERTHRESHOLD:D1 0.5 sets the lower threshold to 0.5 volts.
TRIGGER:A:LOWERTHRESHOLD:D1? might return
TRIGGER:A:LOWERTHRESHOLD:D1 -1.3000 indicating the lower threshold
is set to -1.3 volts.
TRIGger:A:LOWerthreshold{:EXT|:AUX}
Sets or returns the lower threshold for the Auxiliary Input. Used for the following
trigger types: Runt, Slew Rate.
Group Trigger
Syntax TRIGger:A:LOWerthreshold{:EXT|:AUX} {<NR3>|ECL|TTL}
TRIGger:A:LOWerthreshold{:EXT|:AUX}?
Arguments ECL species a preset ECL high level of –1.3V.
TTL species a preset TTL high level of 1.4V.
<NR3> species the threshold level in volts.
TRIGger:A:MODe
Sets or returns the A trigger mode.
Group Trigger
Syntax TRIGger:A:MODe {AUTO|NORMal}
TRIGger:A:MODe?
Related Commands TRIGger:A:LEVel
Arguments AUTO generates a trigger if one is not detected within a specied time period.
NORMal waits for a valid trigger event.
Examples TRIGGER:A:MODE NORMAL species that a valid trigger event must occur before
a trigger is generated.
TRIGGER:A:MODE ? might return :TRIGGER:A:MODE NORMAL indicating that a
valid trigger event must occur before a trigger is generated.
MSO4000 and DPO4000 Series Programmer Manual 2-439