Commands Listed in Alphabetical Order
SAVe:IMAGe:LAYoutSets or returns the layout to use for saved screen images.
Group Save and Recall
Syntax SAVe:IMAGe:LAYout {LANdscape|PORTRait}
SAVe:IMAGe:LAYout?
Arguments LANdscape specifies that screen images are saved in landscape format.
PORTRait specifies that screen images are saved in portrait format.
SAVe:SETUp (No Query Form)Stores the state of the oscilloscope to a specified memory location. You can later
use the *RCL command to restore the oscilloscope to this saved state.
Group Save and Recall
Syntax SAVe:SETUp {<file path>|<NR1>}
Related Commands *RCL,RECAll:SETUp,*SAV
Arguments <file path> is the target location for storing the setup file. <file path> is a
quoted string that defines the file name and path. Input the file path using the
form <drive>:<dir>/<filename>. <extension> and one or <dir>s are optional.
If you do not specify them, the oscilloscope will store the file in the current
working directory. <filename> stands for a filename. (Use of wildcard characters
in filenames is not supported.) Filename extensions are not required but are highly
recommended. For setups, use the extension ".SET".
<NR1> specifies a location for saving the current front-panel setup. The
front-panel setup value ranges from 1 to 10. Using an out-of-range value causes
an execution error. Any settings that have been stored previously at this location
will be overwritten.
Examples SAVE:SETUP 5 saves the current oscilloscope setup in memory location 5.
SAVE:SETUP "TEK00000.SET" saves the current oscilloscope setup in the file
TEK00000.SET in the current working directory.
2-326 MSO4000 and DPO4000 Series Programmer Manual