Commands Listed in Alphabetical Order
Group Acquisition
Syntax ACQuire:NUMAVg <NR1>
ACQuire:NUMAVg?
Related Commands ACQuire:MODe
Arguments <NR1> is the number of waveform acquisitions to average. The range of values is
from 2 to 512 in powers of two.
Examples ACQUIRE:NUMAVG 16 species that 16 waveform averages will be performed
before exponential averaging starts.
ACQUIRE:NUMAVG? might return :ACQUIRE:NUMAVG 64 indicating that there
are 64 acquisitions specied for averaging.
ACQuire:STATE
Starts or stops acquisitions. When state is set to ON or RUN, a new acquisition
will be started. If the last acquisition was a single acquisition sequence, a new
single sequence acquisition will be started. If the last acquisition was continuous,
a new continuous acquisition will be started.
If RUN is issued in the middle of completing a single sequence acquisition (for
example, averaging or enveloping), the acquisition sequence is restarted, and
any accumulated data is discarded. Also, the oscilloscope resets the number of
acquisitions. If the RUN argument is issued while in continuous mode, acquisition
continues.
Group Acquisition
Syntax ACQuire:STATE {OFF|ON|RUN|STOP|<NR1>}
ACQuire:STATE?
Related Commands ACQuire:STOPAfter
Arguments OFF stops acquisitions.
STOP stops acquisitions.
ON starts acquisitions.
RUN starts acquisitions.
2-72 MSO4000 and DPO4000 Series Programmer Manual