Commands Listed in Alphabetical Order
Examples WFMOUTPRE:WFID? might return :WFMOUTPRE:WFID "Ch1, DC
coupling,100.0mVolts/div,500.0µs/div, 1000 points, Sample
mode"
WFMOutpre:XINcr? (Query Only)Returns the horizontal point spacing in units of WFMOutpre:XUNit for the
waveform specified by the DATa:SOUrcecommand. This value corresponds to
the sampling interval. The query command will time out and an error will be
generated if the waveform specified by DATa:SOUrceis not turned on.
Group Waveform Transfer
Syntax WFMOutpre:XINcr?
Related Commands DATa:SOUrce,WFMOutpre:XUNit?
Examples WFMOUTPRE:XINCR? might return :WFMOUTPRE:XINCR 10.0000E-6
indicating that the horizontal sampling interval is 10 µs/point.
WFMOutpre:XUNit? (Query Only)Returns the horizontal units for the waveform specified by the DATa:SOUrce
command. The query command will time out and an error will be generated if the
waveform specified by DATa:SOUrceis not turned on.
Group Waveform Transfer
Syntax WFMOutpre:XUNit?
Related Commands DATa:SOUrce
Examples WFMOUTPRE:XUNIT? might return :WFMOUTPRE:XUNIT "HZ" indicating that
the horizontal units for the waveform are in Hertz.
WFMOutpre:XZEro? (Query Only)Returns the time coordinate of the first point in the outgoing waveform.
2-494 MSO4000 and DPO4000 Series Programmer Manual