Commands Listed in Alphabetical Order
Examples TRIGGER:A:SETHOLD:THRESHOLD:D1 ECL sets the threshold to ECL levels.
TRIGGER:A:SETHOLD:THRESHOLD:D1? might return
TRIGGER:A:SETHOLD:THRESHOLD:D1 -1.3000 indicating the
threshold is set to -1.3 volts.
TRIGger:A{:TRANsition|:RISEFall}? (Query Only)
Returns transition time trigger parameters.
Group Trigger
Syntax TRIGger:A{:TRANsition|:RISEFall}?
Related Commands TRIGger:A:UPPerthreshold:CH<x> ,TRIGger:A:LOWerthreshold:CH<x>
Examples TRIGGER:A::TRANSITION? might return
:TRIGGER:A:TRANSITION:POLARITY POSITIVE;WHEN
SLOWER;DELTATIME 8.0000E-9
TRIGger:A{:TRANsition|:RISEFall}:DELTatime
Sets or returns the delta time used in calculating the transition value for the
transition trigger.
Group Trigger
Syntax TRIGger:A{:TRANsition|:RISEFall}:DELTatime <NR3>
TRIGger:A{:TRANsition|:RISEFall}:DELTatime?
Arguments <NR3> species the delta time, in seconds.
Examples TRIGGER:A:TRANSITION:DELTATIME 15E-6 sets the delta time of the
transition trigger to 15 µs.
TRIGGER:A:TRANSITION:DELTATIME? might return :TRIGGER:A
:TRANSITION:DELTATIME 2.0000E-09 indicating that the delta time of the
transition trigger is set to 2 ns.
2-452 MSO4000 and DPO4000 Series Programmer Manual