Commands Listed in Alphabetical Order
Arguments EDGe is the default trigger. A trigger event occurs when a signal passes through
a specied voltage level in a specied direction and is controlled by the
TRIGger:A:EDGE? commands.
LOGic species that a trigger occurs when specied conditions are met and is
controlledbytheTRIGger:A:LOGIc? commands. This trigger type is equivalent
to the logic trigger and to the setup and hold trigger found in the user interface.
PULSe species that a trigger occurs when a specied pulse is found and is
controlledbytheTRIGger:A:PULse? commands. This trigger type is equivalent
to the pulse width, runt, and rise/fall time triggers found in the user interface.
BUS species that a trigger occurs when a communications signal is found.
Supports CAN, I2C, SPI, and RS-232 communications signals.
VIDeo species that the trigger occurs when a video signal is found.
Examples TRIGGER:A:TYPE EDGE sets the A trigger type to EDGE.
TRIGGER:A:TYPE? might return :TRIGGER:A:TYPE PULSE indicating that the
A trigger type is a pulse trigger.
TRIGger:A:UPPerthreshold:CH<x>
Sets the upper threshold for channel <x>, where x is the channel number. Each
channel can have an independent level. Used only for runt and slew rate trigger
types.
Group Trigger
Syntax TRIGger:A:UPPerthreshold:CH<x> {<NR3>|ECL|TTL}
TRIGger:A:UPPerthreshold:CH<x>?
Arguments <NR3> is the threshold level in volts.
ECL species a preset ECL high level of –1.3V.
TTL species a preset TTL high level of 1.4V.
Examples TRIGGER:A:UPPERTHRESHOLD:CH2 50E-3 sets the upper limit of the pulse
runt trigger to 50 mV for channel 2.
TRIGGER:A:UPPERTHRESHOLD:CH2? might return :TRIGGER:A:
UPPERTHRESHOLD:CH2 1.2000E-01 indicating that the upper limit of the pulse
runt trigger is set to 120 mV.
MSO4000 and DPO4000 Series Programmer Manual 2-455