Commands Listed in Alphabetical Order
Examples TEKSECURE initializes both waveform and setup memories.
This is a program example of how to generate an SRQ when TEKSECURE
completes:
# Bit 0 of the DESE (Device Event Status Enable Register)
# enables OPC to be reported to the SESR (Standard Event
# Status Register)
DESE 255
# Bit 0 of the ESER (Event Status Enable Register)
# enables OPC to be summarized in the ESB (Event Status #
Bit) of the SBR (Status Byte Register)
*ESE 255
# Bit 5 of the SRE (Service Request Enable Register)
enables
# the generation of SRQ when the ESB bit of the SBR
becomes # TRUE
*SRE 32
TEKSECURE;*OPC
When the TEKSECURE operation has completed, the OPC bit of the SESR will be
TRUE and SRQ will have been generated.
TIME
Sets or returns the time that the oscilloscope displays.
Group Miscellaneous
Syntax TIME <QString>
TIME?
Related Commands DATE
Arguments <QString> is a time in the form "hh:mm:ss" where hh refers to a two-digit hour
number, mm refers to a two-digit minute number from 00 to 59, and ss refers to a
two-digit second number from 00 to 59.
Examples TIME "14:00:00" sets the time to exactly 2:00 p.m.
TIME? might return :TIME "14:05:17” indicating the current time is set to 2:05
p.m. and 17 seconds.
MSO4000 and DPO4000 Series Programmer Manual 2-391