Commands Listed in Alphabetical Order
Syntax TRIGger:A:HOLDoff:TIMe <NR3>
TRIGger:A:HOLDoff:TIMe?
Arguments <NR3> species the holdoff time in seconds. The range is from 20 ns through 8.0 s.
Examples TRIGGER:A:HOLDOFF:TIME ? might return :TRIGGER:A:HOLDOFFTIME
1.2000E-06 indicating that the A trigger holdoff time is set to 1.2 µs.
TRIGGER:A:HOLDOFF:TIME 10 sets the A trigger holdoff time to 10 s.
TRIGger:A:LEVel
Sets or returns the trigger level for the A trigger.
Group Trigger
Syntax TRIGger:A:LEVel {ECL|TTL|<NR3>}
TRIGger:A:LEVel?
Arguments ECL species a preset ECL high level of –1.3V.
TTL species a preset TTL high level of 1.4V.
<NR3> species the trigger level in user units (usually volts).
Examples TRIGGER:A:LEVEL? might return :TRIGGER:A:LEVel 1.3000E+00
indicating that the A edge trigger is set to 1.3 V.
TRIGGER:A:LEVEL TTL sets the A edge trigger to TTL high level, which is 1.4 V.
TRIGger:A:LEVel:AUXin
Sets or returns the trigger level for the AUXIN port.
Group Trigger
Syntax TRIGger:A:LEVel:AUXin {<NR3>|ECL|TTL}
TRIGger:A:LEVel:AUXin?
2-428 MSO4000 and DPO4000 Series Programmer Manual