Commands Listed in Alphabetical Order
Group Measurement
Syntax MEASUrement:MEAS<x>:MINImum?
Examples MEASUREMENT:MEAS1:MINIMUM? might return
:MEASUREMENT:MEAS1:MINIMUM 1.75E-09.
MEASUrement:MEAS<x>:SOURCE[1]
Sets or returns the source for all single source measurements and species
the source to measure "from" when taking a delay measurement or phase
measurement. Measurements are specied by <x>, which ranges from 1 through 4.
Group Measurement
Syntax MEASUrement:MEAS<x>:SOURCE[1] {CH<x>|MATH<y>|REF<x>|D<x>}
MEASUrement:MEAS<x>:SOURCE[1]?
Arguments CH<x> is an input channel waveform, where x is the channel number.
MATH<y> is a math waveform, where y is 1.
REF<x> is a reference waveform, where x is the reference channel number.
D<x> is a digital waveform, where x is the digital channel number. Digital
channels are available only on MSO models.
Examples MEASUREMENT:MEAS2:SOURCE1 MATH1 species Math 1 as the measurement
2 source.
MEASUREMENT:MEAS1:SOURCE1? might return
:MEASUREMENT:MEAS1:SOURCE[1] MATH1 indicating that Math1 is the
measurement 2 source.
MEASUrement:MEAS<x>:SOURCE2
Sets or returns the reference source to measure "to" when taking a delay
measurement or phase measurement. Measurements are specied by <x>, which
ranges from 1 through4.
Tip: Source2 measurements only apply to phase and delay measurement types,
which require both a target (Source1) and reference (Source2) source.
2-220 MSO4000 and DPO4000 Series Programmer Manual