Commands Listed in Alphabetical Order
Arguments FALL specifies polarity as the clock falling edge.
RISe specifies polarity as the clock rising edge.
Examples TRIGGER:A:SETHOLD:CLOCK:EDGE RISE specifies the polarity as the clock
rising edge.
TRIGGER:A:SETHOLD:CLOCK:EDGE? might return :TRIGGER:A:SETHOLD:
CLOCK:EDGE RISE indicating that polarity is specified as the clock rising edge.
TRIGger:A:SETHold:CLOCk:SOUrceSets or returns the clock source for the setup and hold triggering.
Group Trigger
Syntax TRIGger:A:SETHold:CLOCk:SOUrce
{CH1|CH2|CH3|CH4|D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|D
1
TRIGger:A:SETHold:CLOCk:SOUrce?
Related Commands TRIGger:A:SETHold:DATa:SOUrce
Arguments CH1–CH4 or D0–D15 specifies the input channel number. D0–D15 is only for
MSO models.
AUX or EXT specifies an external trigger using the Aux Input connector located
on the front panel of the oscilloscope.
Examples TRIGGER:A:SETHOLD:CLOCK:SOURCE CH1 specifies channel 1 as the clock
input for setup and hold input.
TRIGGER:A:SETHOLD:CLOCK:SOURCE? might return :TRIGGER:A:SETHOLD:
CLOCK:SOURCE CH4 indicating that channel 4 is the clock source for the setup
and hold trigger input.
TRIGger:A:SETHold:CLOCk:THResholdSets or returns the clock voltage threshold for the setup and hold trigger.
Group Trigger
MSO4000 and DPO4000 Series Programmer Manual 2-447