Commands Listed in Alphabetical Order
Arguments <NR3> is the vertical offset in digitizing levels.
Examples WFMINPRE:YOFF 50 species that the zero reference point for the incoming
waveform is 50 digitizing levels (2 divisions, for 8-bit data) above the center
ofthedatarange.
WFMINPRE:YOFF? might return :WFMINPRE:YOFF 25 indicating the vertical
position of the incoming waveform in digitizing levels.
WFMInpre:YUNit
Sets or returns the vertical units of the incoming waveform.
Supported units are: %, /Hz, A, A/A, A/V, A/W,A/dB,A/s,AA,AW,AdB,As,
B, Hz, IRE, S/s, V,V/A, V/V, V/W, V/dB, V/s, VV, VW, VdB, Volts,Vs, W,
W/A, W/V,W/W, W/dB, W/s, WA, WV, WW, WdB, Ws, dB, dB/A, dB/V,dB/W,
dB/dB, dBA, dBV,dBW, dBdB, day, degrees, div,hr, min, ohms, percent, s
Group Waveform Transfer
Syntax WFMInpre:YUNit <QString>
WFMInpre:YUNit?
Related Commands WFMOutpre:YUNit?
Arguments <QString> contains a maximum of three alpha characters that represent the
vertical unit of measure for the incoming waveform.
Examples WFMINPRE:YUNIT? might return :WFMINPRE:YUNIT "V" indicating the vertical
units for the incoming waveform are volts.
WFMINPRE:YUNIT "A" species that the vertical units for the incoming waveform
are Amperes.
WFMInpre:YZEro
Sets or returns the vertical offset of the incoming waveform in units specied
by WFMInpre:YUNit. Variations in this number are analogous to changing the
vertical offset of the waveform.
YMUlt, YOFf, and YZEro are used to convert waveform record values to YUNit
values using the following formula (where dl is the data level; curve_in_dl is a
data point in CURVe):
2-486 MSO4000 and DPO4000 Series Programmer Manual