Commands Listed in Alphabetical Order
Table 2-37: DATaand WFMOutpre Parameter Settings (cont.)
WFMOutpre Settings
DATa:ENCdg
Setting :ENCdg :BN_Fmt :BYT_Or :BYT_NR
RPBinary BIN RP MSB 1,2
SRIbinary BIN RI LSB 1,2
SRPbinary BIN RP LSB 1,2
Examples DATA:ENCDG? might return :DATa:ENCDG SRPBINARY for the format of the
outgoing waveform data.
DATA:ENCDG RPBinary sets the data encoding format to be a positive integer
where the most significant byte is transferred first.
DATa:SOUrceSets or returns the location of the waveform data transferred from the oscilloscope
by the CURVe?query.
Group Waveform Transfer
Syntax DATa:SOUrce
{CH1|CH2|CH3|CH4|MATH|REF1|REF2|REF3|REF4|D0|D1|D2|D3|D4|D5|D6|D7|D8
DATa:SOUrce?
Related Commands CURVe
Arguments CH1–CH4 specifies which analog channel data will be transferred from the
oscilloscope to the controller, channels 1 through 4.
MATH specifies that the Math waveform data will be transferred from the
oscilloscope to the controller.
REF1–REF4 specifies which Reference waveform data will be transferred from
the oscilloscope to the controller, waveforms, 1 through 4.
D0–D15 specifies which digital channel data will be transferred from the
oscilloscope to the controller, channels 0 through 15. (MSO models only)
DIGital specifies that the Digital waveform data will be transferred from the
oscilloscope to the controller. (MSO models only)
MSO4000 and DPO4000 Series Programmer Manual 2-149