1023
Marker Function
Detector 1 Dwell Time
This is the time specified by the user to dwell while taking the measurement for detector 1. The
minimum allowed dwell time is based on the current detector If “Off” is selected for detector 1, this key
is grayed out and shows 200 ms.
Detector 2 Dwell Time
This is the time specified by the user to dwell while taking the measurement for detector 2. The
minimum allowed dwell time is based on the current detector. If “Off” is selected for detector 2, this
key is grayed out and shows 200 ms.
Example :CALC:MAM:DET3 QPE
Sets the detector for measure at marker detector 1 to Quasi peak
:CALC:MAM:DET3 OFF
Sets the detector for measure at marker detector 1 to Off
Preset EMI Average
State Saved Saved in instrument state
Readback Text Detector name
Backwards Compatibility SCPI [:SENSe]:EMI:MEASure:DETector:AVERage[:STATe] OFF|ON|0|1|
If sent with On as a parameter, sets detector 3 to EMI Average
If sent with Off as a parameter, sets detector 3 to Off
Initial S/W Revision A.02.00
Key Path Marker Function, Measure at Marker, Detectors
Remote Command See “Detectors” on page 1020.
Example :CALC:MAM:DET:DWEL 400 ms
Sets the dwell time for detector 1 to 400 ms
Preset 200 ms
State Saved Saved in instrument state
Min 1 ms
Max 60 s
Initial S/W Revision A.02.00
Default Unit s
Key Path Marker Function, Measure at Marker, Detectors
Remote Command See “Detectors” on page 1020.