252
System Functions
System
Example: :CAL:TEMP:SOUR:LALL?
Notes: Value is in degrees Centigrade at which the last successful Align Now,
Sourcewas performed on the instrument.
State Saved: No
Initial S/W Revision: A.05.00
Key Path: Visual annotation in the Show Alignment Statistics screen
Mode: All
Remote Command: :CALibration:TIME:LPReselector?
Example: :CAL:TIME:LPR?
Notes: Value is the date and time the last successful Characterize Preselector was
executed. The date is separated from the time by a space character. Returns “”
if no Characterize Preselector has ever been performed on the instrument.
Dependencies: In models that do not include preselectors, this command is not enabled and
any attempt to set or query will yield an error.
State Saved: No
Initial S/W Revision: Prior to A.02.00
Key Path: Visual annotation in the Show Alignment Statistics screen
Mode: All
Remote Command: :CALibration:TEMPerature:LPReselector?
Example: :CAL:TEMP:LPR?
Notes: Value is in degrees Centigrade at which the last successful Characterize
Preselector was executed.
Dependencies: In models that do not include preselectors, this command is not enabled and
any attempt to set or query will yield an error.
State Saved: No
Initial S/W Revision: Prior to A.02.00
Key Path: Visual annotation in the Show Alignment Statistics screen
Mode: All
Remote Command: :CALibration:AUTO:TIME:OFF?
Example: :CAL:AUTO:TIME:OFF?
Notes: Value is the elapsed time, in seconds, since Auto Align has been set to Off or
Off with Alert. The value is 0 if Auto Align is ALL or NORF.
State Saved: No