711
Code Domain
SPAN X Scale
I/Q Error View Evm Window Auto ScalingWhen Auto Scaling is On, and the Restart front-panel key is pressed, this function automatically displays the scale per division and reference value results in the EVM window of the I/Q Error window.Code Domain View Symb Power Window Auto ScalingWhen Auto Scaling is On, and the Restart front-panel key is pressed, this function automatically displays the scale per division and reference value results in the symbol power window of the code domain (quad
Restriction and Notes Upon pressing the Restart front-panel key or Restart menu key under
the Meas Control menu, the Auto Scaling function automatically
determines the scale per division and reference values based on the
measurement results if this parameter is set to On. When you set a
value to either Scale/Div or Ref Value manually, Auto Scaling
automatically changes to Off.
Dependencies/Couplings See Restriction and Notes
Preset ON
State Saved Saved in instrument state.
Range Off|On
Key Path Span X Scale
Mode cdma2000
Remote Command :DISPlay:CDPower:VIEW3:WINDow3:TRACe:X[:SCALe]:COU
Ple 0|1|OFF|ON
:DISPlay:CDPower:VIEW3:WINDow3:TRACe:X[:SCALe]:COU
Ple?
Example DISP:CDP:VIEW3:WIND3:TRAC:X:COUP ON
DISP:CDP:VIEW3:WIND3:TRAC:X:COUP?
Restriction and Notes Upon pressing the Restart front-panel key or Restart menu key under
the Meas Control menu, the Auto Scaling function automatically
determines the scale per division and reference values based on the
measurement results if this parameter is set to On. When you set a
value to either Scale/Div or Ref Value manually, Auto Scaling
automatically changes to Off.
Dependencies/Couplings See Restriction and Notes
Preset ON
State Saved Saved in instrument state.
Range Off|On
Key Path Span X Scale