1377
Trigger
green line will be displayed to indicate the trigger level.
Trig Slo pe
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a
falling edge.
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.
Key Path: Trigger, Baseband I/Q, Q (Demodulated)
Remote Command: :TRIGger[:SEQuence]:QDEMod:LEVel <voltage>
:TRIGger[:SEQuence]:QDEMod:LEVel?
Example: TRIG:QDEM:LEV 0.5 V
Preset: 0.25 V
State Saved: Saved in instrument state
Range: –1 to 1 V
Readback Text: 0.1 of displayed unit (V, mV, etc.)
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Q (Demodulated)
Remote Command: :TRIGger[:SEQuence]:QDEMod:SLOPe POSitive | NEGative
:TRIGger[:SEQuence]:QDEMod:SLOPe?
Example: TRIG:QDEM:SLOP POS
Preset: POSitive
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Q (Demodulated)
Remote Command: :TRIGger[:SEQuence]:QDEMod:DELay <time>
:TRIGger[:SEQuence]:QDEMod:DELay?
:TRIGger[:SEQuence]:QDEMod:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:QDEMod:DELay:STATe?
Example: TRIG:QDEM:DEL 10 ms
TRIG:QDEM:DEL:STAT ON