687
Code Domain
Meas Setup
Meas Interval

Sets the length of the measurement interval in PCG (power control group), 1 PCG = 1536chips.

Meas Offset

Sets the timing offset of measurement interval in PCG (power control group), 1 PCG = 1536 chips

Restriction and Notes IPH – I Phase
QPH – Q Phase
IQC – Combined I and Q Phase
This command is effective when [:SENSe]:RADio:DEVice is set to MS. (This menu
label is blank when [:SENSe]:RADio:DEVice is set to BTS.)
Preset QPH
State Saved Saved in instrument state.
Range I|Q|IQC
Key Path Meas Setup
Unit PCG
Mode cdma2000
Remote Command :CALCulate:CDPower:SWEep:TIME <integer>
:CALCulate:CDPower:SWEep:TIME?
Example CALC:CDP:SWE:TIME 2
CALC:CDP:SWE:TIME?
Restriction and Notes Set the length of the measurement interval that is used. If summation of Meas Interval
and Meas Offset exceeds 32 after changing Meas Interval (or Meas Offset), then Meas
Offset (or Meas Interval) decreases automatically to make the summation 32.
Dependencies/Couplings Max value is dependent on Capture Interval and Meas Offset
Preset 1
State Saved Saved in instrument state.
Min 1
Max Capture Interval
Key Path Meas Setup
Unit PCG
Mode cdma2000