689
Code Domain
Meas Setup
PN OffsetSync Type (BTS only)

Opens a menu that enables you to set the channel type to be used for synchronization.

Unit x 64 chips
Remote Command :CALCulate:CDPower:PNOFfset <integer>
:CALCulate:CDPower:PNOFfset?
Example CALC:CDP:PNOF 5
CALC:CDP:PNOF?
Restriction and Notes Sets value for the pseudo-random noise offset. Different
pseudo-random noise offsets are used for different base stations. By
setting the pseudo-random noise offset to the value that your specific
base station is set to, you get the correct time offset value displayed and
returned back to you when you query READ:CDPower? The
instrument, by default, assumes an offset of 0. So, if you do not use this
command, you have to manually calculate the time offset when the
value is other than 0.
Remote Command Notes You must be in cdma2000 mode to use this command.
Preset 0
State Saved Saved in instrument state.
Min 0
Max 511
Key Path Meas Setup, More
Mode cdma2000
Remote Command [:SENSe]:CDPower:SYNC[:BTS] PICH|DPICh
[:SENSe]:CDPower:SYNC[:BTS]?
Example CDP:SYNC DPIC
CDP:SYNC?
Restriction and Notes This command is effective when [:SENSe]:RADio:DEVice is set to BTS.
Preset PICH
State Saved Saved in instrument state.
Range F-PICH|TxDiv F-PICH
Key Path Meas Setup