System Functions
System
No confirmation is required when Alert is configured through a remote command.
None
With Auto Align Alert set to None the instrument will not signal an alert. This is provided for rare occasions where
you are making a long measurement which cannot tolerate Auto Align interruptions, and must have the ability to
capture a screen image at the end of the measurement without an alert posted to the display. Agilent does not
recommends using this selection in any other circumstances, because of the risk of accuracy performance drifting
well beyond expected levels without the operator being informed.
For front panel operation, confirmation is required to transition into this setting of Alert. The confirmation dialog
is:
No confirmation is required when Alert is configured through a remote command.
Key Path: System, Alignments, Auto Align, Alert
Mode: All
Example: :CAL:AUTO:ALER WEEK
Readback Text: 7 days
Status Bits/OPC dependencies: Bit 14 is set in the Status Questionable Calibration register.
Initial S/W Revision: Prior to A.02.00
Key Path: System, Alignments, Auto Align, Alert