404
ACP Measurement
Meas Setup
Meas Setup

Displays the setup menu for the currently selected measurement. The functions included in the

measurement setup menu include setting the parameters for the carriers, offsets, bandwidths,

measurement methods and types. This menu also allows you to turn noise correction on and off.

Average/Hold Number

Specifies the number of measurement averages used to calculate the measurement result. The average

will be displayed at the end of each sweep. After the specified number of average counts, the average

mode (termination control) setting determines the average action.

Key Path Front-panel key
Initial S/W Revision Prior to A.02.00
Key Path Meas Setup
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command [:SENSe]:ACPower:AVERage:COUNt <integer>
[:SENSe]:ACPower:AVERage:COUNt?
[:SENSe]:ACPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:AVERage[:STATe]?
Example ACP:AVER:COUN 250
ACP:AVER:COUN?
ACP:AVER OFF
ACP:AVER?
Notes You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Preset 10
ON
State Saved Saved in instrument state.
Min 1
Max 1000
Backwards Compatibility SCPI [:SENSe]:ACPR:AVERage:COUNt
[:SENSe]:MCPower:AVERage:COUNt (PSA Power Suite, PSA W-CDMA,
PSA cdma2000)
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00