944
Input/Output
corrections that are marked as ON (see “Correction On/Off” on page 934) are used.
Delete All Corrections
Erases all correction values for all 4 Amplitude Correction sets.
When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to
delete all corrections. Press ESC or Cancel to close this dialog.” The deletion is only performed if you
press OK or Enter.
Remote Correction Data Set Commands
Set (Replace) Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas.
The values sent in the command will totally replace all existing correction points in the specified set.
An Ampcor array can contain 2000 points maximum.
Key Path: Input/Output, Corrections
Remote Command: [:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Example: SENS:CORR:CSET:ALL OFF
This command makes sure that no amplitude corrections are applied,
regardless of their individual on/off settings.
Preset: Not affected by Preset. Set to OFF by Restore Input/Output Defaults
State Saved: Saved in instrument state.
Initial S/W Revision: A.02.00
Key Path: Input/Output, Corrections
Remote Command: [:SENSe]:CORRection:CSET:ALL:DELete
Example: CORR:CSET:ALL:DEL
Initial S/W Revision: A.02.00
Remote Command: [:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA <freq>,
<ampl>, . . .
[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA?
Example: CORR:CSET1:DATA 10000000,–1.0,20000000,1.0
This defines two correction points at (10 MHz, –1.0 dB) and (20 MHz, 1.0
dB) for correction set 1.
Preset: Empty after Restore Input/Output Defaults. Survives a shutdown or restart of
test set application (including a power cycle).