1308
Trace/Detector
Notes The four states of this 1-of-N actually set two variables, Update and Display,
to their four possible combinations:
Trace On: Update and Display both On
View: Update Off and Display On
Blank: Update Off and Display Off
Background: Update On, Display Off
See tables below for detail on the SCPI to control these two variables.
Couplings Selecting a trace type (Clear Write, Trace Average, Max Hold, Min Hold) for
a trace (pressing the key or sending the equivalent SCPI command) puts the
trace in Trace On (Update On and Display On), even if that trace type was
already selected.
Selecting a detector for a trace (pressing the key or sending a
[:SENS]:DET:TRAC command) puts the trace in Trace On (Update On and
Display On), even if that detector was already selected.
Selecting a math mode other than Off for a trace (pressing the key or sending
the equivalent SCPI command) puts the trace in Trace On (Update On and
Display On), even if that math mode was already selected.
Loading a trace from a file puts that trace in View regardless of the state it was
in when it was saved; as does being the target of a Copy or a participant in an
Exchange.
Initial S/W Revision Prior to A.02.00