249
System Functions
System
Show Alignment Statistics
Shows alignment information you can use to ensure that the instrument is operating in a specific manner.
The Show Alignment Statistics screen is where you can view time and temperature information.
Values which are displayed are only updated when the Show Alignment Statistics screen is invoked, they
are not updated while the Show Alignment Statistics screen is being displayed. The remote commands
that access this information obtain current values.
An example of the Show Alignment Statistics screen would be similar to:
A successful Align Now, RF will set the Last Align RF temperature to the current temperature, and reset
the Last Align RF time. A successful Align Now, All or Align Now, All but RF will set the Last Align
Now All temperature to the current temperature, and reset the Last Align Now All time. A successful
Align Now, All will also reset the Last Align RF items if the RF portion of the Align Now succeeded.
Notes: Only available in: XOBT
:CAL:SOUR? Initiates an Alignment and returns 0 if successful
:CAL:SOUR? Initiates an Alignment and returns 1 if failed
Couplings: Initializes the time for the Last Align Source Now, All Time.
Records the temperature for the Last Align Source Now, All Temperature.
Status Bits/OPC dependencies: Bits TODO may be set in the Status Questionable Calibration register
Initial S/W Revision: A.05.00
Key Path: System, Alignments
Mode: All