416
ACP Measurement
Meas Setup
Offset/Limits
Accesses a menu of functions that contains Offset, Offset Freq/Offset To Edge, Offset Integ BW, Upper
Offset Limit and Lower Offset parameters.
Select Offset
Selects the offset to configure.
Offset Freq
This parameter determines the frequency difference between the center of the main channel and the
center of the carrier.
Each Offset Freq state value is entered individually by selecting the desired carrier on the carrier menu
key using the up down arrows, RPG or numeric keypad. Then enter the Offset Freq State using the Offset
Frequency key.
The list contains up to six (6) entries, depending on the mode selected, for offset frequencies. Each offset
frequency in the list corresponds to a reference bandwidth in the bandwidth list.
An offset frequency of zero turns the display of the measurement for that offset off, but the measurement
is still made and reported. You can turn off (not use) specific offsets with the
Preset 0.22
C2K: No
DTMB: 0.05
State Saved Saved in instrument state.
Min 0.01
Max 1.0
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00
Key Path Meas Setup
Initial S/W Revision Prior to A.02.00
Key Path Meas Setup, Offset/Limits, Limits
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE
Preset A
State Saved Saved in instrument state.
Range Offset A|Offset B|Offset C|Offset D|Offset E|Offset F
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00