697
Code Domain
Peak Search
Marker Delta
Sets the control mode for the selected marker to Delta mode.
Continuous Peak Search
Turns Continuous Peak Search on or off. When Continuous Peak Search is on, a peak search is
automatically performed for the selected marker after each sweep. The rules for finding the peak are
exactly the same as for Peak Search, including the peak criteria rules. If no valid peak is found, the “No
Peak Found” warning is generated after each sweep. If a valid peak is found, the message “Peak Found”
is displayed after each sweep.
Pk-Pk Search
Finds and displays the amplitude and frequency (or time, if in zero span) differences between the highest
and lowest y-axis value.
Remote Command :CALCulate:CDPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|1
2:MAXimum:LEFT
Example CALC:CDP:MARK2:MAX:LEFT
Key Path Peak Search
Key Path Peak Search
Mode cdma2000
Remote Command :CALCulate:CDPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|1
2:CPEak[:STATe] ON|OFF|1|0
:CALCulate:CDPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|1
2:CPEak[:STATe]?
Example CALC:CDP:MARK:CPE ON
CALC:CDP:MARK:CPE?
Preset OFF
State Saved Saved in instrument state.
Range Off|On
Key Path Peak Search
Mode cdma2000
Remote Command :CALCulate:CDPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|1
2:PTPeak
Example CALC:CDP:MARK:PTP
Restriction and Notes Turns on the Marker Δ active function.